Warnings and precautions

Whenever the

symbol appears on the NucleoCounter

®

NC-

3000™ instrument, it

indicates that the manual must be consulted for precautions and warnings.

Power and cables

Use the shielded USB cable supplied with the NucleoCounter

®

NC-

3000™ to ensure

that appropriate EMI classification is maintained for the intended environment.

The USB interface connector of the NucleoCounter

®

NC-

3000™ must only be

connected to SELV circuits. External computing devices connected to the USB

interface connector of the NucleoCounter

®

NC-

3000™ has to comply with the

standards, UL 1950 and IEC/EN 60950.

Table 1. Description of the NucleoCounter

®

USB interface connector

NucleoCounter

®

NC-

3000™ is powered by an external 24VDC power supply. For

safe use, please follow the instructions for connecting the power supply.

The NucleoCounter

®

NC-

3000™ shall only be used with an external power

supply purchased at ChemoMetec A/S.

The detachable DC power supply cord set and appliance inlet of the external

power supply are considered as the disconnecting device.

The mains supply cord and plug of the external power supply shall comply with

any national regulations.

1

In normal operation mode (refers to Pin no 4)

Pin no.

Name

Maximum

Voltage level

1

1

+5V

+5 VDC

2

D-

+3.5 VDC

3

D+

+3.5 VDC

4

DGND

0 VDC

Metal

enclosure

Shield

(connected

to DGND)

0 VDC