Actuator movements, Nserting and removing the slide, 3 inserting and – ChemoMetec NC-3000 User Manual

Page 32: Removing the slide

Advertising
background image

5 Operation of the

991-3001 Rev. 1.3

June-2013

16

5.2.1 Actuator movements

When a Cassette is analyzed after Run is pressed, the piston actuator will move the

piston rod down through the cylinder of the Cassette. Hereby, the sample material is

transported to the Cassette measurement chamber. The form and dimensions of the

flow channels facilitate effective mixing of the sample. Sensors monitor sample flow.

When the sample has been correctly loaded into the measurement chamber, the

piston is stopped.

5.3

Inserting and removing the slide

To insert the slide:

1. Eject the sample tray.

2. Hold the slide at the end opposite to the arrows.

3. Place the slide onto the sample tray, by giving it a gentle push with the

thumb.

4. Insert the sample tray and/or Run

4

the analysis.

After analysis the Slide is removed from the sample tray and disposed of as

biological waste according to national or regional laws or regulations regarding the

nature of the mixture it contains .

Figure 10. Illustration of how to hold and place the slide on the sample tray

4

The Run operation will always insert the sample tray.

Advertising