5 Operation of the

991-3001 Rev. 1.3

June-2013

17

5.4

Result presentation

Depending on the type of analysis performed the result will be presented on the PC

within a few seconds and the fluorescent image(s) will be stored on the hard disk .

Typical process time is from 20 seconds to 2 minutes for multi image and multi color

analysis.

5.5

Protection of the optical system

The lid covers the sample tray. This lid is designed to protect the area from dust and

other potential contaminants, and to keep external light from interfering with the

fluorescent image recording. The lid is spring loaded so it will automatically close

during insertion of the sample tray.

Unless inserting or ejecting a sample media or cleaning the sample tray, the sample

tray should be inserted at all times to protect the optical system of the

NucleoCounter

®

NC-3000

™. After operation, shut down the NucleoView NC-3000™

software first, so the sample tray will be inserted in case left ejected. Then the

NucleoCounter

®

NC-

3000™ can be turned off by removing the DC power plug or

disconnecting the wall outlet.