7 Maintenance of NucleoCounter

®

NC-3000

991-3001 Rev. 1.3

June-2013

21

7 Maintenance of NucleoCounter

®

NC-3000

7.1

Cleaning

7.1.1 Instrument enclosure.

Depending on the environment in which the NucleoCounter

®

NC-

3000™ is operated,

it is suggested that regular cleaning of the enclosure is carried out. W hen cleaning

the enclosure it is recommended to use a soft moist cloth and gently wipe the

surface. Any contamination, which does not come off immediately, should be rubbed

gently with a cloth wetted with mild detergent. Never use organic solvents or

aggressive detergents to clean the exterior of the NucleoCounter

®

NC-

3000™ as

this might damage the surface. Cleaning of the enclosure on a daily basis can be

performed by using a mild det ergent.

7.1.2 Sample tray

When cleaning the sample tray, great care must be taken against introducing any

liquid or dust into the NucleoCounter

®

NC-3000

™. Any liquid that enters the interior

of the NucleoCounter

®

NC-

3000™ can damage the optical parts and thus

compromise the quality of the analyses.

The black surface of the sample tray is anodized aluminum and can therefore be

wiped off with a clean, dry and dust free cloth that reduces fiber contamination of

the sample tray.

The sample tray and the optical parts inside the NucleoCounter

®

NC-

3000™ should

be properly protected against dust and other contaminants. Therefore, great care

must be taken to ensure that the sample tray is inserted when the instrument is not

in use. If the black surface of the sample tray becomes contaminated it should

immediately be cleaned with a clean, dry and dust free cloth.

7.1.3 Removal of dust particles

Two optical windows are placed in the sample tray. Please ensure that these are

kept clean at all times. In case cleaning of a window is needed then refer to the

Cleaning kit NucleoCounter

®

and follow the instructions here (ChemoMetec Part no:

911-0013).