23

3

C

HAPTER

3: C

ALIBRATION

The 60 Series indicators can be calibrated several ways. The following
method is a quick calibration procedure and assumes that the necessary
parameters are selected before the actual calibration is performed (i.e. full
scale value, graduation size, etc.). Your local GSE distributor will normally
set up these parameters.

Q

UICK

C

ALIBRATION



A certified weight is required to perform a calibration. Serious
inaccuracies could result from using non-certified standards for calibration.


To access calibration on the Model 460:

• From the Weigh Mode press

[ZERO]+[SELECT]

[ZERO] [PRINT] [UNITS] [TARE/ENTER]


To access calibration on all other 60 Series

indicators:

• From the Weigh Mode press

[100] [SELECT] [5 4 3 2 1] [ID] [ENTER]

The controller will briefly prompt you with QUICK CAL! And then New
Zero?
. If multiple scales are enabled, the controller will prompt you with
Keyin Scl #. Key in the scale number being calibrated and press [ENTER].
Use the [PRINT] and [UNITS] keys to enter the scale number on the Model
460. Refer to Table 2: 460 Key Functions on page 8. After the scale number
is entered the New Zero? prompt will be displayed.

Press the [SELECT] key to toggle through the available calibration
selections, and [ENTER] when the desired calibration method is displayed.
Press [ENTER] to begin calibration when the desired calibration selection is
displayed. Pressing the [CLR] key at any point in the calibration routine will
back you up one step.