6 0 S

E R I E S

I

N D I C A T O R S

58

06 Tare>F.S.!

Tare entry was greater than full scale. Most
likely the entered tare value was incorrect.

07 Tare < 0 !

Negative tare attempted, but not allowed per
P162. For auto-tares, the GROSS Weight must
be greater than zero unless P162 is changed to
allow negative tares.

08 CheckConn.

Displayed if the signal into the A/D is +/- 2
times the Full Scale signal. This is effectively
taken into consideration when the information
sent to the microprocessor from the A/D is +/-
twice the allowable F.S. reading. ie. P110 F.S.
= 100. Error message will be displayed at +/-
208 taking into consideration the 4% overload.

S

ETUP

M

ODE

E

RROR

M

ESSAGES

10 Entry>Max!

An entry was made which had more characters
than allowed.

11 WRONGCODE!

The incorrect access code was entered, thus
preventing changes. In order to access the
Setup Mode, either the proper code must be
entered or [ENTER] must be pressed alone (to
view selections without making changes).

12 No Mods!

The Setup Mode is being accessed, but changes
are prevented.

13 OutOfRange

An entry made for a selection was beyond the
range of valid choices.

14 Must Keyin

The choice for the current parameter must be
keyed in.

15 Size>3999!

The size of a Custom Transmit setup has
exceeded the limit.

16 CHECK JUMPR

A programming operation was attempted when
the program jumper is installed. Installation of
this jumper prohibits programming changes.