61

34 RES> 25K!

The current combination of capacity P110 and
increment P111 result in a resolution greater
than 25,000 graduations. This is simply a
warning in case this was not intended.

35 RES>100K!

The current combination of capacity P110 and
increment P111 result in a resolution greater
than 100,000 graduations. This is not allowed
and as soon as any key is pressed the instrument
will jump back into the Setup Mode to
parameter P110 to verify the settings.

36 RES< 100!

The current combination of capacity P110 and
increment P111 result in a resolution less than
100 graduations. This is simply a warning in
case this was not intended.

37 RES< 1 !!

The current combination of capacity P110 and
increment P111 result in a resolution less than 1
graduation (i.e. the increment is greater than
capacity). This is not allowed and as soon as
any key is pressed the instrument will jump back
into the Setup Mode to parameter P110 to verify
the settings.

G

ENERAL

E

RROR

M

ESSAGES

99 Can't Set!

An attempt to enter a value for a parameter
which is not field changeable, such as the serial
numbers or the audit trail counter results in this
message.

- - Cksumerror

Upon each power-up, the indicator tests the
integrity of its EPROM. If the result is not
correct this message is displayed and the
Indicator is not usable. Verify that the EPROM
(U13) is installed properly (no bent pins).