P205 will appear in the “Custom Setup” list for each receive port enabled.
For example, if the receive mode is enabled for all four ports, the list will
display

P205—

,

P205˜

,

P205™

, and

P205š

.

P240 – Minimum Transmit Width

A weight value that cannot be displayed due to the 6-digit limitation of the
standard VF display may not be printed. To ensure this is not possible, P240
must be set to a width of not greater than 7 (6 digits and a decimal point).
NTEP must also be enabled at P440. Any weight value that exceeds the
minimum width specified will be printed as dashes "-------".

P440 – NTEP Enable

P440 appears in the “Custom Setup” list if disabled. Refer to Enabling
NTEP Operation
on page 46 for possible implications.

P9990 – Macro Instance Selection

P9990 appears in the “Custom Setup” list if at least one macro is configured.
Macro operation should be checked to verify its conformance to all
regulations.

A

DDITIONAL

C

ONFORMANCE

C

ONSIDERATIONS

Several parameters must be considered on an individual basis as their
configuration may vary with different applications. These parameters
include, but are not limited to those listed in Table 5.

Table 5: Additional Conformance Parameters

P

ARAMETER

D

ESCRIPTION

C

OMMENT

P110

Full Scale Capacity

Verify proper scale capacity.

P111

Division Size

Verify allowable scale divisions.

P112

Zero Track

Verify required selection.

P114

Motion

Verify required selection.

P118 Zero

Range

Verify required selection.

P212

Print Stability

Verify required selection.

P126 Æ P130

Multi-Range

Verify proper configuration.

P151 Æ P154

Custom Units

Verify name and conversion factor.

P600 Æ P646

Rename Parameters

Verify acceptable names.