Kistler-Morse KM Microcell Bolt-On User Manual

Page 19

Advertising
background image

Note

Ground the cable shield only at the signal
processor.

INSTALLING INSULATION FOR
OUTDOOR VESSELS (OPTIONAL)

The sun affects the performance of an outdoor,
bolt-on sensor system. The sun’s radiation
heats the support metal unevenly, producing
stresses in the supports that are unrelated to the
weight of material in the vessel. The Microcell™
system minimizes errors associate with
sun-induced stressed in several ways.
A. Microcell™ sets and instrumentation of all

support legs allow the system to subtract

bending stresses resulting from uneven heating

of supports.

B. Microcell™ rosette arrays, where

applicable, allow the system to subtract

tensile/compressive stresses resulting from the

heating of supports.

This configuration of the Microcell™ system
minimizes errors associated with sun-induced
stresses. However, if Microcells™ are installed
on the legs between braces (See Figure 3-6),
insulatio on each of the adjacent braces is required
for best performance. This “brace wrap”
insulation increases system accuracy by further
reducing sun-induced stresses.

INSULATION ORDER AND INSTALLATION
EQUIPMENT

The following are included with the insulation
order (quantities are dependent of the number
of braces):

Brace wrap, 60 in by 85 in (1.5m by 2.2m)

Tie wraps

The following are used for installation:

Flexible tape measure

Heavy-duty knife

INSTALLING BRACE WRAP

1. See Figure 3-17. Using a flexible tape

measure, measure and record the wrap width

required. allowing for a minimum 2 in (51mm)

overlap.

2. See Figures 3-17 ad 3-18. Lay the wrap on a

flat surface. Mark and cut it at the distance

from Step 1.

3. See Figure 3-19. The goal is to cover most

of the brace with wrap. Covering the brace

where it crosses another brace in the middle

is unnecessary. Depanding on the brace lenth,

multiple sections of wrap may be required,

with each section overlapping the one below

it by a minimum of 2 in (51mm). Measure and

record the space available for each section of

wrap. If the space is more than 60 in (1.5m),

skip Step 4 and proceed to Step 5.

Note

If a junction box is mounted within the area to
be covered by wrap, cut the wrap so it does not
cover the junction box.

4. From the top edge, measure and mark the

wrap at the distance from Step 3. Cut the

wrap where marked.

5. Position the wrap, starting at the bottom of the

brace. Wrap it around the brace, overlapping

the ends as shown in Figure 3-17. Fasten the

wrap to the brace with four tie wraps.

6. Repeat Steps 2 through 5 for additional

sections of wrap. Overlap each section of

wrap by a minimum of 2 in (51mm).

Figure 3-17. Wrap on Various Shapes.

15

Advertising