Checking equipment – Kistler-Morse KM Microcell Bolt-On User Manual

Page 7

Advertising
background image

MICROCELL™ INSTALLATION EQUIPMENT

Tape measure
Marking pen
Kistler-Morse

®

Test Meter

Kistler-Morse

®

Microcell™ Sensor Drill Template

Drill motor
Tapping fluid
Tap handle
Disk grinder, 4½” (114 mm) or larger, or belt
grinder
Sandpaper (coarse and fine)
Degreaser (isopropyl alcohol or acetone)
Level
Caulking gun

9

/

64

” hex T-handle driver

Digital Multimeter (DMM)
Tape (electrical or masking)

Note

If the Microcell™ sets will be install by
Kistler-Morse

®

, the service technician provided

by Kistler-Morse

®

will bring this equipment on

site with the tool kit. If the Microcell™ sets will
be installed by the customer, the purchase
of a Kistler-Morse

®

Test Meter is highly

recommended to simplify the installation.

JUNCTION BOX AND FIELD WIRING
EQUIPMENT

Drill motor
#29 drill bit
#8-32, 2-flute, spiral-point tap
Tap handle
Tapping fluid

9

/

64

” Allen wrench

#8-32 socket head cap screws
#8 flat washers (

3

/

16

” inner diameter,

7

/

16

” outer

diameter)
Belden™ 8791, 18 gauge, 3-conductor shielded
interconnect cable or equivalent (for up to
1,000 ft (305m) length)
Belden™ 8618, 16 gauge, 3-conductor shielded
interconnect cable or equivalent (for 1,000 ft
to 2,000 ft (305m to 610m) length)
Conduit and fittings or cable tray
Caulking gun
Sikaflex 1A polyurethane sealant or Dow
Corning RTV 738 or RTV 739

CAUTION

Only use Sikaflex 1A polyurethane sealant or
Dow Corning RTV 738 or RTV 739.
Other
sealants may contain acetic acid, which is
harmful to sensors and electronics.

CHECKING EQUIPMENT

CAUTION

Handle Microcell™ sets with care. Dropping,
striking, etc. can damage the Microcells™.

VISUAL CHECK

Visually inspect all equipment in the order,
including Microcells™, junction boxes,
Installation Kit, and insulation (if provided), to verify
they were not damaged during shipment. If any
item was damaged, contact Kistler-Morse

®

immediately for a replacement.

FUNCTIONAL CHECK

Perform a functional check of the Microcells™
before installation to verify they were not damaged
during shipment. Two methods of performing the
check are described below.

TESTING WITH A KISTLER-MORSE

®

TEST METER
The Kistler-Morse

®

Test Meter (Figure 2-1) is

designed specifically to test Kistler-Morse

®

sensors. If you do not have a test meter,
disregard this section and proceed to TESTING
WITH A DIGITAL MULTIMETER (DMM).

Note

The test meter display indicates a low battery or
behaves erratically when the batteries are weak.
When this occurs, replace the batteries before
testing.
1. See Figure 2-1. Connect the Microcell™

red, white, and black wires to the

corresponding test meter terminals. Place

the Microcell™ on a stable surface.

2. Turn on the power to the test meter and set

the Simulate/Test Switch to the Test

position. Verify the no-load output is

between +25mV and -25mV.

3. Repeat Steps 1 and 2 for each Microcell™.

If the no-load output for any Microcell™ is

outside these specifications:

3

Advertising