Manual calibration – Kistler-Morse KM Microcell Bolt-On User Manual

Page 32

Advertising
background image

REMOVING A KNOWN QUANTITY
OF MATERIAL

See Figure 5-1.
1. Record the current live load.
2. Input HI Span: HI Span

= current live load.

3. Remove known quantity of material to the

vessel. Ensure all material has stopped

moving before proceeding.

4. Input LO Span: LO Span = HI Span - Removed

Weight.

Example
You are using Microcells™ to monitor a vessel. The vessel
contains 110,000 lb of material and can hold a maximum of
200,000 lb. You plan to remove 60,000 lb of material (which is
greater than 25% of 200,000 lb).
Following the Live Load Calibration procedure:
1. Current live load = 110,000 lb
2. HI Span = current live load = 110,000 lb
3. Remove 60,000 lb of material.
4. LO Span = HI Span - Removed Weight

= 110,000 lb - 60,000 lb

= 50,000 lb

MANUAL CALIBRATION

Note

The Kistler-Morse

®

SVS-2000™ signal

processor performs a manual calibration
automatically with Quick Config.

See Figure 5-2.
1. Refer to the signal processor manual to

determine how to obtain the A/D converter

sensitivity, expressed in Counts/mV. Record

this value.

2. Record the Microcell™ sensitivity (S).

Sensitivity for Microcells™ on legs and beams

are shown in Table 5-1.

3. Refer to the Application Data Form for the

vessel (contact Kistler-Morse

®

for an additional

copy if needed). Record the maximum live

load and the stress.

4. Record the current live load in the vessel.

5. Calculate the Manual Calibration values:

Scale Factor Weight = maximum live load

Scale Factor Counts = S * Counts/mV * Stress

Zero_Cal = current live load

6. Refer to the signal processor manual to input

the calibration values.

Example 1. Microcells™ on Vertical Legs.
You are using 3” Microcells™ in rosette arrays on vertical
column legs. The vessel has four W10x39 carbon steel legs
and no braces. The vessel currently contains 50,000 lb of
material and can hold a maximum of 200,000 lb.
Following the procedure:
1. Counts/mV = 699.05 (from signal processor)
2. S = 0.045 mV/psi (From Table 5-1, for legs with rosette

array)

3. From the Application Data Form, the maximum live load

is 200,000 lb. The stress is 4348 psi.

4. Current live load = 50,000 lb
5. Calculate the values for the calibration:

Scale Factor Weight = Maximum live load = 200,000 lb

Scale Factor Counts = S * Counts/mV * Stress

= 0.045 mV/psi * 699.05 Counts/mV * 4348 psi

= 136,776 Counts

Zero_Cal = current live load = 50,000 lb

Example 2. Microcells™ on Beams.
You are using 3” Microcells™ on beams. The vessel has four
W10x39 carbon steel horizontal beams and four W10 x 39
carbon steel diagonal beams. The Microcells™ are on the
horizontal beams only. The vessel currently contains
50,000 lb of material and can hold a maximum of 150,000 lb.
Following the procedure:
1. Counts/mV = 699.05 (from signal processor)
2. S = 0.070 mV/psi (From Table 5-1, for beams)
3. From the Application Data Form, the maximum live load

is 150,000 lb. The stress is 5929 psi.

4. Current live load = 50,000 lb
5. Calculate the values for the calibration:

Scale Factor Weight = Maximum live load = 150,000 lb

Scale Factor Counts = S * Counts/mV * Stress

= 0.070 mV/psi * 699.05 Counts/mV * 5929 psi

= 290,127 Counts

Zero_Cal = current live load = 50,000 lb

Figure 5-2. Manual Calibration Line.

28

Advertising