Error messages 15 – ALTANA Micro-Haze Plus User Manual

Page 46

Advertising
background image

45

Do not attempt to repair the unit yourself. If your measuring unit should
experience a malfunction, our costumer service will be glad to help you
as quickly as possible.

Fluctuations of
the measured
values

Was the same spe-
cimen position used
for all the measure-
ments?

No: Check how high
the deviations are on
the specimen itself.

Error Messages

15

Yes: If applicable,
check the calibration
using the auxiliary
standard.

Calibration incorrect:
Recalibrate, possibly
clean the standard
(chap. 7).

Calibration correct:
Is the specimen sur-
face completely
smooth and does the
measuring unit make
good contact with the
specimen?

No: In this case, very
large deviations are
possible (chap. 10).

Yes: The unit is
defective.

Please contact our
customer service.