4.2.3 Load the Machine:

Refer to Appendix A (“Cleaning Cannulated Instruments”) for suggestions about the care of your
laparoscopic instruments.

1. Instruments may be loaded into tray with the tray located in the Irrigator tank, inside the machine, or

resting above the bath. (Use the “T” handle on the tray to position tray above bath)

2. Dismantle the instruments and insert the nozzle end of the cannulated instrument (standard sizes 2, 5

and 10mm) into the connector provided.

3. With the ports vertical, place the connector, with instrument loaded, into the port.

4. Gently lay the instruments down and rest them on the moveable plastic rack. Adjust the rack to

support the instruments. The other parts of the instruments may be placed on the bottom of the tray
or in the instrument holders provided.

5. The unused ports should be left in vertical position (when the Irrigator function is turned off to that

port).

6. Lower the tray into the cleaning bath. (Ensure the “T” handle is on the right hand side).

7. Ensure that the spigot in the bath docks onto the instrument tray.

4.2.4 Start Cleaning:

1. Close the lid and press “Clean”

2. The machine will perform a cleaning cycle according to its program.

3. The machine will perform a short de-gas, and then commence cleaning.

4. The following table explains the various parameters of the clean function, which are displayed

during cleaning and are programmable to suit your requirements.

Cleaning
Activity

Display

Parameters set by manufacturer
(Settings)

Explanation

Clean No.

C-01

10

Current Cleaning cycle
number

Ultrasonics

P-03

3

Displays power level

Time
Remaining

09.48

10 mins

Time remaining to complete
current clean cycle shown in
minutes and seconds.

Irrigator

Red Light
on
keypad

5
(one rest period = 5 secs)

Light indicates if function is
on or off (irrigate for 5 secs,
rest for 25 secs, and repeat)

Table 2: Displays during the cleaning process

4.2.5 To Pause Cleaning:

1. Lift the lid to pause the cleaning cycle.

2. This feature allows for any adjustments the instruments may require.

4.2.6 To Stop Clean Cycle:

1. Press “Clean” again during cycle. This will stop cycle and clear current clean.

2. To start new cycle press “Clean” again.

4.2.7 Completion of Single Clean Cycle:

1. The machine, will emit 5 beeps and display ‘- - - -‘.