Appendix D: Recommended

S-2800/S-2800MRS disinfection protocol

It is Soniclean’s protocol to disinfect every irrigator supplied for trial at the end of every trial and
immediately prior to commencing the next trial. Soniclean or their distributors/agents will ensure this is
undertaken unless the trial organisation prefers to conduct the disinfection themselves.

The protocol to be followed:

1. Drain irrigator and turn off at power point.

2. Apply protective gloves (disposable non sterile latex gloves are suitable).

3. Remove tray; thoroughly wipe down all visible surfaces of the irrigator including the lid, tank, all

outside surfaces, hoses and attachments with disinfectant solution

1

.

4. Remove detergent hose and replace with disinfectant hose connected to bottle with slightly more

than required amount of disinfectant

1.2

. Insert tray.

5. Turn irrigator on at power point whilst holding down the program key (this will allow you to re-

program the parameters).

6. Change soap % parameter to required amount of disinfectant. Ensure irrigate function on.

7. Press “clean”. The irrigator will now add appropriate amount of disinfectant to tank and in doing so

will disinfect the internal plumbing fittings as well as irrigator ports.

8. Once clean is complete, disconnect disinfectant bottle and replace with bottle filled with clean

water.

9. Press “shift” and “refill” keys, hold for three seconds. This will cause irrigator to empty of water and

disinfectant and then to fill with water. This will rinse disinfectant out of the unit.

10. Repeat steps 5 and reprogram soap parameter to factory setting (i.e. 1.2)

11. Dry all external surfaces and pack unit ready for transport.

___________________________________________________________________

Note regarding Disinfectant solutions: The recommendation is to use a TGA approved medical grade
disinfectant which is recommended for use in hospital settings for disinfecting horizontal and vertical
working surfaces.

Soniclean currently uses a product called LCC (supplied by Dominant Chemicals). For disinfecting the

S-2800 Irrigator Operating Manual

Page 21

© 2013-2014 Soniclean Pty Ltd

Version 20

March 2014

KEY TO PROGRAMMING DISPLAY
PARAMETER

DISPLAY

MACHINE

MIN

MAX

FACTORY
SETTINGS

Clean time

Ct

mins

1

99

10

Ultrasonic power level

PO

1

3

3

No. of cleans between
drains

Cn

0

99

10

Detergent concentration

SP

percentage

0.0

5.0

0.5

Start up degassing time

G1

mins

0=Off

99

10

Degassing time prior to
each clean

G2

mins

0=Off

99

2

Test time

Tt

sec

0=Off

99

30

Irrigate default

Id

0=Off

1=On

1

Irrigate rest periods

Ir

1 rest period
= 5 seconds

5 period
(5 x 5 = 25
seconds)

50

5