Appendix C: Recommended S2800/S2800MRS preventative protocol

Soniclean recommended the following protocol to be carried out (daily and weekly by your authorities;
annually by your service agent) to ensure the machine is performing at the optimum performance.

DAILY

Remove and clean strainer

Ensure strainer is clean.

Drain machine at end of day/session

Ensure the contaminated water is not stored in tank

Check cleaning efficacy

Visual examination of all load items.

IMPORTANT NOTE:

If the water stored in the machine for a long duration, water might trapped in the level sensor tube. This
might potentially cause the error in machine level sensor, and hence, the machine will not stop
automatically during water overflowing.

WEEKLY

Safety checks

Check condition of lid hinges and tray

Lid interlock test

Ensure safety to operator

Drainage test

Ensure free-flowing drainage

Detergent dosing test

Ensure the detergent can flow into the tank

ANNUALLY

Verification of calibration e.g. plumbing system, detergent intake, level sensor, electronic boxes,
MRS

Completion of all validation & performance tests

A service reminder sticker and Test Certificate will be provided after each calibration.

Where to obtain service? Return your machine to Soniclean for a full overhaul and service.

Preventative maintenance contract recommendations - please contact Soniclean for more
information.

Re-ordering information for SONIC 1 detergent.

S-2800 Irrigator Operating Manual