F.1 for technician – Soniclean S2800 User Manual

Page 29

Advertising
background image

F.1 For Technician:

Please tick relevant boxes. If an item is not applicable write “N/A”.

Plumbing:

i)

Screw supplied braided hose on to the machine and inlet mains water supply (hand tighten only).

ii)

Screw and hand tighten supplied grey sink hose onto the machine and insert into drain.

iii) Ensure that the drain hose falls evenly away from the machine to the drain

Electrical:

i)

Wall power socket 240 VAC 50-60 Hz rated at 10 amps.

ii)

Wall power socket located within 600mm from the machine.

iii) Power supply circuit shall not be shared with life threatening or life support electrical systems.
iv) Power supply circuit shall not be shared with computer electrical system or other excessively large

power devices, e.g. steriliser/washer disinfectors.

v)

Earth leakage protection shall be provided by user.

Machine Reporting System: (MRS only)

i)

LAN connection; take the LAN provided cable and push the black plug into the MRS socket provided
on the Irrigator. Plug the other end (RJ45) into the LAN wall outlet.

ii)

Contact your IT department and hand them the Machine Reporting Manual. The appropriate
information for connecting the machine to the Local Area Network is found on page 10 “Network
Hardware and Configuration” of the Machine Reporting System Manual.

iii) PC Connection: take the PC cable provided and push the black plug into the MRS socket provided on

the Irrigator. Plug the other end (RJ45) into the appropriate outlet on the PC.

iv) Temperature Sensor: Push the ambient temperature sensor probe into the temperature sensor socket

found on the right side of the machine.

Printer:

i)

Plug the machine into mains supply and turn on at the switch.

ii)

Plug the printer cable into the machine and the printer

iii) Press the feed button to progress the paper.

Detergent Bottle:

i)

Connect detergent hose to left hand side socket of machine.

ii)

Place filtered end of detergent hose into container.

iii) Ensure there is sufficient detergent in bottle at all times. Explain this to customer that level
must be checked regularly.
iv) Please note Chemical Type:_______________________(Sonic 1) Dosage Setup:_________(0.5%)

Machine Location:

i)

Place machine on flat and horizontal surface (maximum 2º tilt).

Caution: Explain to the customer the following

Only use designated cables for accessing data from the Irrigator.
The cable marked ‘PC’ should be used with a laptop computer.
The cable marked ‘LAN’ should be used for connection to a LAN wall outlet.

S-2800 Irrigator Operating Manual

Page 25

© 2013-2014 Soniclean Pty Ltd

Version 20

March 2014

Advertising