F.2 S2800 Customer Installation Requirements

IMPORTANT NOTE: The customer shall ensure the following requirements are met prior to installation.
This is essential for the machine to perform at optimum level and to fulfil Australian Standards.

Environmental:

i)

Use water with ambient temperature

ii)

A tap or inlet must be dedicated to the irrigator.

iii) Clean water (meet AS/NZS 4187:2003) must be used for cleaning surgical instruments.
iv) A drain outlet must be dedicated to the irrigator.

Plumbing:

i)

Maximum water pressure 150 psi, minimum water pressure 50 psi.

ii)

Flow rate 10 litres/minute (minimum) to 20 litres/minute (maximum).

iii) This is a High Hazard application and the inlet mains supply shall be fitted with an approved

Reduced Pressure Zone Device and a line filter. (Please check with your local Water Authority for the
correct installation information for your area.)

iv) Mains inlet connection with ½” male BSP fitting within 600mm from the machine.
v)

Do not use any aeration devices in the mains water supply.

vi) Outlet pipe of 25mm diameter for waste water within 600mm from the machine.

Machine Reporting System: (MRS models only):

i)

Local Area Network (LAN) outlet 1000mm from the machine.

ii)

When bottle is empty replace with the same brand of detergent. It is important the identical detergent
is used, as this is critical to the total cleaning process. If other detergent is used the machine shall
require re-programming.

Machine Location:

i)

Place the machine on a flat and horizontal surface. (Maximum 2º tilt)

ii)

Position at operator’s convenient height. (The top of the lid at elbow height)

iii) The bench top shall be strong enough to support the fully loaded machine. (approximately 60

kilograms)

iv) The unit requires 590mmx 880mm of bench space
v)

Allow minimum 100mm clearance at the rear to enable plumbing access. This may be reduced by
fitting elbows to both the inlet and the outlet.

vi) Allow 1000mm clearance above unit to allow the lid to open fully.

S-2800 Irrigator Operating Manual