1.4 Programmable Features

When you receive your Soniclean Irrigator it will be set to function automatically. The parameters are set in
the factory. The features described below may be altered as you require. The list includes the display you will
see on the machine when you are programming. To alter these parameters refer to section 6, Program Mode.

Parameter display name

Descriptions

Ct

Length of cleaning times

PO

Ultrasonic power levels

Cn

Clean time. This determines the number of cleaning cycles your machine will
perform before automatically draining. To refill, press SHIFT and hold down REFILL
for three seconds.

SP

Detergent concentration

G1

Start up degassing time

G2

Degassing time prior to each clean, (pre-clean degas)

tt

Test time. Set time for foil test

Id

Irrigate default. Re-programs the machine to suspend the irrigator function.

Ir

Irrigate rest periods. Determines the On/Off cycles of the irrigator function.

1.5 Display Messages

The following messages are displayed on the keyboard at various times during normal operation. The display
cycles through the items marked with “*” while cleaning.

Display

Explanation

- - - -

Together with 5 beeps, this indicates that a process has been completed. The operator
does not need to wait for the beeps to complete before starting another process.

04.58*

Shows the time remaining of the clean or degas function which is currently operating.

C-01*

Indicates the current cleaning cycle number.

P-03*

Indicates power level selected, e.g. P-02, P-03, etc.

dGAS

The machine is de-gassing.

drn

The machine is draining.

drnf

Bottom of the wash tank is flushed with clean water, and then drained.

FILL

The machine is filling the wash tank.

OFF

The machine has completed a drain or other operation and needs to be turned off at
power point.

END

End of final clean.

Lid

When the lid is in the open position, a flashing “Lid” message is displayed, and all
operations are paused. When the lid is lowered again the paused operations will
recommence from the point of the pause.

tESt

When the lid is open to maintain ultrasonic (foil/UPM) performance test

S-2800 Irrigator Operating Manual

Page 4

© Soniclean Pty Ltd

Version 20

March 2014