☞ Condensate and deposits in the pressure transducer affect the measuring result.

☞ In case of deposits, aggressive or condensable media, install a gas washing bottle

before the pressure transducer if necessary.

☞ In order to avoid malfunction it is important to position the pressure transducer in

the vacuum line so as to avoid flow of condensate towards the pressure trans-
ducer.

☞ Clean pressure transducer if necessary, see section ”Cleaning and assembling

components”.

Max. permitted pressure at the pressure transducer: 1.5 bar (ab-
solute).
☞ The display flashes at a pressure higher than 1100 mbar.

☞ Comply with max. permitted pressure.

☞ Inside a vacuum system where evaporation occurs, e. g.

rotary evaporator, the vacuum is not uniform, e. g. a cold
condenser acts as pump or the vacuum in the pipeline is
lower than in the system. Therefore carefully choose posi-
tion where to connect the gauge head.

gauge head
VSK 5

Notes on operation

Assembling the enclosed components (PC 2008/2010/2012 VARIO):

inlet

➨ Assemble separator with centring ring and clamping ring to

the inlet of the pump.

fixing screw

➨ Connect the cable of the gauge head to the female connec-

tor at the rear side of the vacuum controller.

➨ Connect the gauge head to the vacuum system by using a

small flange connection or a hose connection.

☞ The device is adjusted together with the gauge head at the

factory. If the gauge head is replaced a readjustment is rec-
ommended.

➨ Connect cable from the frequency converter of the pump to

the controller (rear side): trip line and power supply of the
frequency converter.

female connector to
connect gauge head
VSK 5

➨ Assemble controller CVC 2000

II

with fixing screw at the con-

troller support.

The separator at the inlet prevents droplets and particles from
entering the pump.
☞ Lifetimes of diaphragms and valves are enhanced.

☞ Improves vacuum in case of condensation.