☞ Remove union nut at the condenser and remove hose from

the inlet of the condenser.

☞ Pull off old overpressure safety relief device and install new

one.

☞ Connect hose to the inlet of the exhaust waste vapour con-

denser and tighten union nut.

Overpressure safety relief device ......................... 63 88 21
(at the exhaust waste vapour condenser)

Cleaning and assembling components

Cleaning the gauge head VSK 5

The gauge itself is maintenance-free.

If the vacuum system is contaminated (oil, particles, etc.), contamination of the pressure transducer will
influence the accuracy of measurement.

Do not clean the pressure transducer using a spiky or hard tool!

➨ Fill the gauge head chamber with a solvent (e. g. benzene) and allow sufficient cleaning time. Observe

all regulations concerning usage and disposal of solvents!

➨ Drain the solvent and dispose of in accordance with regulations, repeat cleaning if necessary.

➨ Rinse the gauge head chamber several times with alcohol in order to remove all solvent residues.

➨ Allow the pressure transducer to dry.

➨ Readjust the pressure transducer if necessary.

Spare parts:

Exhaust waste vapour condenser .......................................................................................... 69 99 75
Collecting flask 1000 ml
.......................................................................................................... 63 88 77
Separator (at the inlet)
............................................................................................................ 69 99 80

Calibration in the factory

Control of measuring equipment

The VACUUBRAND DKD calibration laboratory is accredited by the Physikalisch-Technische
Bundesanstalt (PTB; German national institute for science and technology and the highest technical authority
of the Federal Republic of Germany for the field of meteorology and certain sectors of safety engineering)
for the measurable variable pressure in the pressure range from 10

-3

mbar to 1000 mbar in accordance

with the general criteria for the operation of testing laboratories defined in the DIN EN ISO/IEC 17025:2000
series of standards.
Calibration in the VACUUBRAND calibration laboratory:
-

To meet the requirements of the DIN ISO 9000ff and 10012 series of standards regarding the calibra-
tion of inspection, measuring and test equipment at specified intervals.

-

To document that the vacuum gauges calibrated are traceable to national standards of the PTB.