General view basic setup VACUU•LAN

setting process parameters

➨ lower pressure value (condition for automatic shut down)

➨ upper pressure value (condition for restart)

➨ time for automatic shut down

setting the basic setup

basic setup VACUU•LAN

start: pump down

The pump (and the coolant valve if applicable) is switched

off if the pressure is below the lower pressure value for
longer than the time for automatic shut down.

The pump starts automatically if the pressure is higher

than the upper set point or at great pressure increase.

connect and

configurate

cooling

water valve

yes

no

connect and

configurate

venting valve

yes

no

switching coolant

valve?

switching venting

valve?