To chose a pressure value for automatic switching off is
only possible if automatic switching off is activated (in ba-
sic setup vacuum control with automatic).
➨ Press key MODE simultaneously with ▼ (arrow down).

☞ When the key is pressed shortly, the current pressure for

automatic switching off is displayed for one second (”AE”,
i.e. automatic switching off, is preselected at the factory).
With a second tip within one second or continuous press-
ing:

➨ Select pressure for automatic switching off by using keys ▲

or

▼ (pressure selectable in a range from 1 to 1060 mbar

or ”AE”; ”AE”, i. e. automatic switching off, is active if a pres-
sure value < 1mbar is selected).

☞ The pressure is stored as new pressure for automatic switch-

ing off when the key is released.

Automatic switching off

MODE

30

mbar

Changing to basic setup continuous pumping

➨ Press key MODE simultaneously with ▲ (arrow up).

☞ Switching to continuous pumping is only possible, if vacuum

control is not activated (after pressing key START/STOP).

☞ In the basic setup continuous pumping the pumping speed

can be changed by presetting the motor speed.

☞ Pressing key MODE a second time: switching back to basic

setup vacuum control (only possible when pumping down).

MODE