appropriate mode, vacuum control with or without automatic, continuous pump-
ing with preset pumping speed or VACUU•LAN
: The components of the chemistry
vacuum system (coolant valve, venting valve) have to be preset once only. The mode
as from last operation and the preselected values (e. g. for pressure, speed or time for
switching off) are stored. In case of similar operation conditions it is possible to start
immediately, if the preselections are chosen appropriately.

Working with the controller

How to configurate the controller

Vacuum control:
☞ Position control on preset pressure value.

☞ The pumping speed is adapted to the occurring gas/vapour,

so that the preset pressure value is kept constant.

☞ Operating in the equilibrium of the process by continuous

adaptation of the controller, no abrupt operation of valves.

☞ Vacuum supply according to the requirements for lock cham-

bers or pumping systems not compressing against atmos-
pheric pressure, e. g. wide range turbomolecular pumps or
cryo pumps.

Continuous pumping with preset pumping speed:
☞ Simple suction, sucking liquids by siphon, drying, sucking

off excesses, etc.

☞ Suction with adjustable pumping speed.

☞ Pumping down until the ultimate vacuum of the pump is

achieved (e. g. to dry a system).

☞ System controlled by an interface: temporally or pressure

dependant variable pumping speed to optimise the process
conditions (e. g. to dry sensitive substances).

☞ Pumping with pressure selection ”HI”: Optimum pumping

speed at any pressure.

Vacuum control in automatic mode:
☞ Searching the boiling point and adjusting the set point auto-

matically if the boiling point changes (due to temperature
change, change of ingredients of solutions).

☞ Adjusting the set point automatically after manual setting of

the boiling point (additional automatic mode in basic setup
vacuum control).

☞ Automatic switching off after complete evaporation of the

solvent, if preselected.

In all basic setups:
☞ If cooling water and/or venting valve have been connected

and preselected:

☞ Process-adapted operation of the coolant valve.

☞ Key-driven venting or adjustment of the set point by venting

if process control is active.

-

the components of the VACUUBRAND chemistry vacuum system which are con-
nected to the system.

-

the specific user and/or process requirements.

The vacuum controller CVC 2000

II

offers three basic setups depending on

The controller CVC 2000

II

can be adapted to the specific application by choosing the