VACUUBRAND MV 10C VARIO User Manual

Page 26

Advertising
background image

page 26 of 51

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999073 / 15/12/2009

Automatic mode

Switching on the automatic mode
➨ Press key MODE.

☞ ”Auto” is displayed.

☞ After starting process control, the pressure is adapted to

the process automatically: The controller determines the
boiling point and adapts if the boiling pressure changes.

☞ If automatic mode is switched on while process control is

active, the process pressure is adapted.

Setting the maximum speed:
Set the value for the speed by using key

▲ or ▼ (effective

only if the process pressure is adapted):
☞ To activate set mode: Press key shortly. The current speed

is displayed for one second.

☞ With a second tip within one second or continuous press-

ing:

☞ Set speed. The new set point is stored when the key is re-

leased.

☞ The reduced maximum speed restricts the evaporation rate

in case of low boiling solvents and a carry over of sub-
stances.

☞ The utilisation of a condenser or a cold trap at the inlet can

be optimised.

If selecting ”HI” (recommended) for motor speed, the

maximum speed for automatic adaptation of the motor
speed is restricted to a value corresponding to the sys-
tem size.

10.0

MODE

MODE

MODE

If automatic switching off was preselected and the controller
has determined that evaporation is completely finished:
☞ ”Auto” is flashing.

☞ The pump is switched off. If the controller measures a pres-

sure > 1060 mbar, the pressure is reduced to 1060 mbar to
avoid overpressure.

☞ The controller is still active, the control must be finished by

pressing key START/STOP or by venting (the coolant valve
will not be closed until then and until the after-running time
has elapsed).

To use the advantages of the automatic mode optimally, switch
on ”Auto” before starting the process.
☞ Switching on automatic mode: Press key MODE.

☞ The controller determines the operating pressure.

☞ When activating the automatic mode if process control is

active (after having pressed key START/STOP) the actual
pressure is stored as set point and adjusted automatically.

Switching off the automatic mode
➨ Press key MODE.

☞ Switching off the automatic mode is possible at any time.

☞ ”Auto” disappears.

☞ While vacuum control is active the actual pressure is stored

as new set point.