Read commands – VACUUBRAND MV 10C VARIO User Manual

Page 36

Advertising
background image

page 36 of 51

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999073 / 15/12/2009

Read commands

Function

Command Response

Description

actual pressure

IN_PV_1

XXXX mbar or

unit according to preselections

XXXX Torr or
XXXX hPa

actual frequency

IN_PV_2

XX.X Hz

device set

IN_CFG

XXXXX

preselections

0: remote operation off
1: remote operation on
0: no automatic switch off
1: automatic switch off
0: no venting valve
1: venting valve
0: no coolant valve
1: coolant valve
0: VACUU•LAN
1: continuous pumping
2: vacuum control without automatic
3: vacuum control with automatic

malfunction

IN_ERR

XXXX

1: last command to interface incorrect
1: malfunction pressure transducer
1: overpressure
1: malfunction frequency converter

status of

IN_STAT

XXXX

process control

00: VACUU•LAN:

inactive

01:

pumping down, actual pressure
> preset pressure

02:

pumping down, time for automatic
switching off is running

03:

system is switched off

10: continuous pumping: inactive
11:

active

20: vacuum control:

inactive

21:

actual pressure
> preset pressure

22:

actual pressure =
preset pressure (+/- 1mbar)

23:

actual pressure
< preset pressure

30: automatic:

inactive

31:

determining boiling point

32:

adjusting boiling point

33:

system is switched off

0: venting valve not driven (closed)
1: venting valve driven (open)
0: coolant valve not driven (closed)
1: coolant valve driven (open)