Setting the lower pressure value (condition for automatic shut
down):
➨ Press p▲ or p▼.

☞ ”Set p” and the lower pressure value are displayed for approx.

1s.

☞ With a second tip within one second or continuous press-

ing:

➨ Set the lower pressure value by pressing key p▲ or p▼.

☞ The lower pressure value is stored as the new set point

when the key is released.

Note: It is recommended to select a pressure value which is
approx. 10 mbar higher than the ultimate pressure value achiev-
able in the system.

Setting the process parameters

Setting the time for automatic shut down:
➨ Press key ▲ or ▼.

☞ ”Set”, the clock symbol and the time for automatic shut down

are displayed for approx. 1 s.

☞ With a second tip within one second or continuous press-

ing:

➨ Press key ▲ or ▼ to set the value for the time for shut down

(1 - 200 min).

☞ The value for the shut down time is stored as the new set

point when the key is released.

☞ If the actual pressure falls below the lower set point, the

time for automatic shut down starts to run. The time for au-
tomatic shut down is reset if the actual pressure rises higher
than the lower set point (with a hysteresis of 3 mbar).

The process control stops when the time for shut down has
elapsed. The vacuum pump is switched off and the coolant valve
is closed.
☞ Pump starts again in case of great pressure increase.

☞ If the actual pressure is rising above the upper set point, the

pump starts again (corresponds to starting by pressing key
START/STOP).

☞ The process control can be stopped manually at any time

by pressing key START/STOP.

Setting the upper pressure value (condition for restart):
➨ Keep key MODE pressed and press key ▲ additionally.

☞ ”Set p”, ”arrow up” and the upper pressure value are dis-

played for approx. one second.

247

mbar

67

mbar

MODE

67

mbar

p

015

☞ With a second tip within one second or continuous press-

ing:

➨ Set the upper pressure value by pressing key ▲ or ▼.

☞ The upper pressure value is stored as the new set point

when the key is released.