Technical data, Protection, “technical data”, pg. 24) – VACUUBRAND MZ 2C NT VARIO without controller (o.C.) User Manual

Page 24

Advertising
background image

page 24 of 76

Technical data

General technical data valid for all pumps/pumping units

Maximum permissible inlet pressure

(absolute)

psi

(bar)

16

(1.1)

Maximum permissible outlet pressure

(absolute)

psi

(bar)

16

(1.1)

Maximum pressure difference between

inlet and outlet

psi

(bar)

16

(1.1)

Maximum permissible pressure (abso-

lute) at gas ballast valve

psi

(bar)

17.5

(1.2)

Permissible ambient temperature

storage / operation

°F

(°C)

14 to 140 / 50 to 104

(-10 to +60 / +10 to +40)

Permissible relative atmospheric mois-

ture during operation (no condensation)

%

30 to 85

Maximum permissible installation altitude

above mean sea level

ft

(m)

6500

(2000)

Rated motor power

hp

(kW)

0.77

(0.575)

No-load speed 50/60 Hz

rpm

30 - 2400

Maximum permissible range of supply

voltage ( ±10% )

Attention: Observe specifications of rat-

ing plate!

100-120 V~ 50/60 Hz,

200-230 V~ 50/60 Hz

Motor protection

thermal cutout, manual reset

Coolant connection

(waste vapor condenser)

hose nozzle for tubing

I.D. 1/4” - 5/16”

(hose nozzle DN 6-8 mm)

Maximum permissible pressure of

coolant at waste vapor condenser

psi

(bar)

87 (absolute)

(6 (absolute))

Permissible range of coolant temperature

(waste vapor condenser)

°F

(°C)

5 to 68

(-15 to +20)

The pump achieves its ultimate pumping speed and ultimate vacuum

only at operating temperature (after approx. 15 min.).

Advertising