Fittings and tubing of the different pump models – VACUUBRAND MZ 2C NT VARIO without controller (o.C.) User Manual

Page 64

Advertising
background image

page 64 of 76

Outlet side

MV 10C VARIO-B

PC 3010 VARIO

MD 12C VARIO-B

PC 3012 VARIO

Inlet side

Pumps VARIO-B /

Pumping units PC 301x VARIO

Fittings and tubing of the different pump models:

Disassemble the housing covers (A) to

check the diaphragms and valves.

Unscrew four Allen screws (R) with a 5mm

wide Allen key. Remove the housing cover

with housing cover insert (A) together with

the head cover (D), the valves (B) and the

O-rings (C) (head cover with valves and O-

rings will remain within the housing cover).

+

Never remove parts by using a pointed or sharp-edged tool (e.g., screw

driver). We recommend to use a rubber mallet or compressed air (to

be blown carefully into port).

Remove the head cover (D) carefully from the housing cover. Note the

position and orientation of the valves (B) and remove them.

Replace valves and/or O-rings if necessary.

+

Use petroleum ether or industrial solvent to remove deposits. Do not

inhale vapors.

R

Advertising