Disconnect hose connection between pump and exhaust waste vapor

condenser. Loosen union nut at exhaust waste vapor condenser and

pull hose out of the inlet of the condenser.

Collecting flask at the outlet: Remove joint clip, remove flask and drain

condensate. Observe applicable regulations when disposing of conden-

sate which may be contaminated by pumped chemicals.

Disassembling pump from base plate:

For maintenance, lay the pumping unit on its side with the rating plate

of the pump and the exhaust waste vapor condenser at the top. Sup-

port the pumping unit appropriately.

Remove the hex nuts underneath the base plate with an open ended

wrench (w/f 10) from the feet of the pump (pay attention to serrated

washers). Take the pump off the base plate.

Detach separator (18) from pump inlet

(open clamping ring).

+

Avoid the release of pollutants.

Observe applicable regulations when dis-

posing of condensate which may be con-

taminated by pumped chemicals.

Loosen the union nuts (N) of the hose con-

nections at the pump heads with an open-

ended wrench (w/f 17).

Turn the fittings (Q) with an open-ended

wrench (w/f 14, at outlet w/f 16) to detach

the hoses from the pump heads (1/4 of a

turn at maximum).

+

Do not remove the elbow fittings from the

pump heads; during reassembly a leak may

result.

Q

Disconnect controller from pump (VACUU

BUS cable).

Disassembling the pump from the pumping unit

(PC 3010/3012 VARIO)