The replacement of the diaphragm and the replacement of the valves can

be carried out separately.
+

To replace the valves, remove the head covers of one side of the pump

along with the assembled valve heads and fittings.

+

To maintain the diaphragms, the valve heads and the fittings need not

be disassembled. The head covers can be removed along with the as-

sembled valve heads and fittings.

+

Service only one side of the pump at a time to avoid the mixing of

parts.