Shutdown & storage

The pump can be switched off under vacuum.

Short-term:

Has the pump been exposed to condensate?

- Allow the pump to continue to run at atmospheric pres-

sure for a few minutes.

Has the pump been exposed to media which may damage

the pump materials or form

deposits?

- Check and clean pump heads if necessary.

Long-term:

- Take measures as described above regarding short-

term shutdown.

- Separate the pump from the application.

- Close inlet and outlet ports (e.g., with transport caps).

- Close the gas ballast valve.

- Drain catchpots.

- Store the pump under dry conditions.