VACUUBRAND MZ 2C NT VARIO without controller (o.C.) User Manual

Page 54

Advertising
background image

page 54 of 76

Disassemble the exhaust vapor condenser

EK:

Loosen the union nut at the inlet of the ex-

haust vapor condenser.

Remove the 4 screws affixing the counter

holder of the exhaust vapor condenser with

a Torx driver T10. Remove the exhaust va-

por condenser. In doing so pull the PTFE

hose out of the inlet of the exhaust vapor

condenser.

For maintenance, lay the pump / pumping unit on its side with the

pump heads to be maintained at the top. Support the pump / pumping

unit appropriately.

PC 3002/3003/3004 VARIO

Remove catchpots (14) at inlet and outlet

(see “Use and operation”, pg. 36).

Remove the 4 screws affixing the head cov-

er cowling (W) with a Torx driver T20. Pay

attention to the washers under the screws

and remove.

Pull off head cover cowling (W) carefully. Do

not tilt.

W

Detach the coupling of the connection tube (V)

to the other side of the pump at the valve head

(O).

PC 3002/3003/3004:

Detach the coupling of the connection tube (V)

to the other side of the pump as well as the

hose connection to the inlet/outlet of the vacu-

um system at the valve head (O).

V

O

U