Pumping unit PC 3004 VARIO with exhaust waste va-

por condenser Peltronic at the outlet:

The exhaust waste vapor condenser Peltronic

is separately packed and has to be assembled

at the pumping unit prior to use:

Affix EK Peltronic with the mounting angle

to the pumping unit (1).

Slip the PTFE tube (connecting tube be-

tween pump outlet and inlet of Peltronic)

onto the hose connection of the Peltronic

and fasten with union nut (2).

Check all hose connections prior to starting operation of

the cooling system.

Secure coolant hoses at the hose nozzles (e.g., with hose

clamps) to prevent their accidentally slipping off.

Prevent the discharge of dangerous gases and vapors

to the surrounding atmosphere. If appropriate, connect

the exhaust line to a suitable treatment system.

+

Never block the gas outlet ((8) hose nozzle for tubing

I.D. 3/8” (10 mm)). The exhaust hose has always to be

unobstructed and without back pressure to enable an

unhindered discharge of gases and protect the pump

valves from damage.

Note: Install the hoses of the cooling system in a way

to avoid the flow / dripping of condensed water onto the

pumping unit (especially cables and electronic parts,

see also IP degree of protection, “Technical data”, pg.

24).

• Ensure that the

coolant outlet tubing is always unob-

structed and that it cannot get blocked.

• Maximum permissible coolant pressure at the exhaust

waste vapor condenser: 87 psi (6 bar) absolute. Outlet

flow must always be unhindered.