Due to the high compression ratio, the pump may gen-

erate overpressure at the outlet. Check pressure com-

patibility with system components (e.g., exhaust pipe-

line or exhaust valve) at the outlet.

+

Do not permit any

uncontrolled pressurizing. Make

sure that the exhaust tubing cannot become blocked

to avoid a

risk of bursting!

+

Keep the electrical power cord away from heated sur-

faces.

• Provide a firm, level platform for the equipment. Check

that the system which you are going to evacuate is

mechanically stable. Check that all fittings are secure.

Ensure a stable position of the pump without any me-

chanical contact other than the pump feet.

• Comply with

maximum permissible pressures at in-

let and outlet and with

maximum permissible pres-

sure differences between inlet and outlet. See sec-

tion “Technical data”, pg. 24”. Do not operate the

pump with overpressure at the inlet.

• Avoid overpressure of more than 17.5 psi absolute (1.2

bar absolute) in the event that inert gas is connected to

the pump, to the gas ballast or to a venting valve.

Note: Flexible elements will shrink when evacuated.

• Connect hoses gas tight at inlet and outlet of the pump.

• Ensure that no foreign objects can be drawn into the

pump.

• Check the power source and the pump’s rating plate

to be sure that the power source and the equipment

match in voltage, phase, and frequency.

• Ensure that the

coolant outlet pipe is always free and

that it cannot get blocked. If installing an optional cool-

ant valve, it must always be in the supply line of the

exhaust waste vapor condenser.