In case of condensation, Notice – VACUUBRAND MV 10C NT + EK User Manual

Page 41

Advertising
background image

page 41 of 92

In case of condensation

In case of

condensation: Check the liquid lev-

el in the catchpots during operation. Check the

liquid level in the catchpots regularly. Do not

allow the catchpots to overfill. Drain catchpots

in time to avoid overflow. Install a level sensor

(see “Accessories / Spare parts”, pg. 43) for

monitoring, if necessary (VACUUBRAND con-

troller CVC 3000 or VNC 2 is required).

The maximum liquid level is at approximately

80% of the total filling level to avoid problems

when removing the catchpots.

Drain catchpots.

+

Important: Comply with regulations when

disposing of solvents/condensates. Recy-

cle if possible; purify if contaminated.

Removing the catchpots:

Catchpot at outlet:

Remove joint clip. Remove catchpot and drain

condensate.

Catchpot at inlet:

Admit air or inert gas (via the pump inlet) to

restore atmospheric pressure in the catchpot

before attempting removal. Remove joint clip.

Remove catchpot and drain condensate.

Reattach drained catchpots.

NOTICE

Advertising
This manual is related to the following products: