Prevent the discharge of dangerous gases and vapors

to the surrounding atmosphere. If appropriate, connect

the exhaust line to a suitable treatment system.

Check all hose connections prior to starting

operation of the cooling system.

Secure coolant hoses at the hose nozzles

(e.g., with hose clamps) to prevent their acci-

dentally slipping off.

EX

IN

+

Never block the gas outlet (hose nozzle for

tubing I.D. 3/8” (10 mm)). The exhaust hose

has always to be unobstructed and with-

out back pressure to enable an unhindered

discharge of gases and protect the pump

valves from damage.

• Only install the optional coolant valve in the

supply line of the exhaust waste vapor con-

denser.

Note: Install the hoses of the cooling system

in a way to avoid the flow / dripping of con-

densed water onto the pumping unit (espe-

cially cables and electronic parts, see also

IP degree of protection, “Technical data”,

pg. 24).

• Ensure that the

coolant outlet tubing is

always unobstructed and that it cannot get

blocked.

• Maximum permissible coolant pressure at

the exhaust waste vapor condenser: 87 psi

(6 bar) absolute. Outlet flow must always be

unhindered.

• Comply with the maximum permissible cool-

ant pressures of additional components in

the coolant circuit (e.g., coolant valve).

• Avoid overpressure in the coolant circuit

(e.g., caused by blocked or squeezed cool-

ant hoses).