Electrical connection, During operation, Notice – VACUUBRAND MV 10C NT + EK User Manual

Page 37: Electrical connection during operation

Advertising
background image

page 37 of 92

Permissible range of coolant temperature at the exhaust

waste vapor condenser: 5°F to 68°F (-15°C to +20°C)

Check hose connections prior to starting operation of the

cooling system.

Check coolant hoses regularly during operation.

NOTICE

Plug in the power cord.

• Check the power source and the pump’s

rating plate to be sure that the power source

and the equipment match in voltage, phase,

and frequency.

Switch the pump on.

O

➨ I

Electrical connection

During operation

Vent and dispose of

potentially dangerous gases or

vapors at the outlet of the pump appropriately.

+

Due to the high compression ratio, the pump might

generate overpressure at the outlet. Check pressure

compatibility with system components (e.g., exhaust

tubing or exhaust valve) at the outlet. Ensure that the

pump outlet is neither blocked nor restricted.

+

Maximum ambient temperature: 104 °F

(40 °C)

Check the maximum temperatures, if install-

ing the pump in a cabinet or a housing.

Make sure ventilation is adequate, especial-

ly if the ambient temperature is elevated.

max.

104°F

(40°C)