Make sure ventilation is adequate to maintain recommend-

ed operating temperature. Keep a minimum distance of

2 in (5 cm) between the cooling fan and surrounding items

(e.g., housing, walls, etc.), or else install an external au-

tomatic ventilation system. Do not place the pump on soft

surfaces (e.g., rubber foam) during operation. This may

reduce the air circulation in the pump housing and affect

the fan’s cooling effect. Check fan regularly for dust/dirt.

Clean fan guard grill if necessary to avoid a reduction of

ventilation.

Use only hoses at the inlet and outlet of the pump with

an inner diameter at least as large as the diameter of the

pump’s tubing (to avoid overpressure at the outlet, and

reduction of pumping speed at the inlet).

Allow the equipment to equilibrate to ambient temperature

if you bring it from cold environment into a room prior to

operation. Notice if there is water condensation on cold

surfaces.

Secure coolant hoses at the hose nozzles (e.g., with hose

clamp) to prevent their accidental slipping.

Comply with all

applicable and relevant safety require-

ments (regulations and guidelines). Implement the re-

quired actions and adopt suitable safety measures.

Ambient conditions

Do not reach for this product if it has fallen into liquid.

There is a risk of deadly electrical shock. Unplug the

system immediately.

+

Do not use this product in an area where it can fall or

be pulled into water or other liquids.

• Adopt suitable measures in case of differences from

recommended conditions, e.g., using the equipment

NOTICE