Safety during operation – VACUUBRAND MV 10C NT + EK User Manual

Page 18

Advertising
background image

page 18 of 92

protective action for pump and environment.

You must take suitable precautions to prevent any for-

mation of explosive mixtures in the pump chamber or

at the outlet of the pump. In case, e.g., of a diaphragm

failure, mechanically generated sparks, hot surfaces

or static electricity may ignite these mixtures. Use inert

gas for gas ballast or venting, if necessary.

Drain appropriately or otherwise remove any poten-

tially explosive mixtures at the outlet of the pump, or

dilute them to non-explosive concentrations.

Never operate this pump if it has a damaged cord or

plug.

+

If the pump is not working properly, has been dropped

or has fallen into water, contact your pump service pro-

vider.

+

Prevent any part of the human body from coming into

contact with vacuum.

+

If there is an exhaust isolation valve, make sure that you

cannot operate the equipment with the valve closed to

avoid a

risk of bursting!

+

Always provide a free and pressureless exhaust outlet

to avoid damage to pump valves and risk of bursting.

+

Check the overpressure safety relief device at the ex-

haust waste vapor condenser at appropriate intervals.

+

Comply with applicable regulations when disposing of

chemicals. Take into consideration that chemicals may

be contaminated. Take adequate precautions to pro-

tect people from the effects of dangerous substances

(chemicals, thermal decomposition products of fluoro-

elastomers). Use appropriate protective clothing and

safety goggles.

+

Interruption of the pump (e.g., due to power failure),

failure of connected components or of parts of the sup-

Advertising