Connecting the outlet, Connecting the outlet notice, Notice – VACUUBRAND MV 10C NT + EK User Manual

Page 34

Advertising
background image

page 34 of 92

Outlet via hose nozzle DN 15 mm or via hose

nozzle DN 10 mm (exhaust waste vapor con-

denser EK).

In case, install a small flange connection in-

stead (thread G1/2”, see “Accessories / Spare

parts”, pg. 43).

outlet

If dangerous or polluting fluids could be released at

the outlet, install an appropriate system to catch and

dispose of those fluids.

Connecting the outlet

NOTICE

EK

+

Connect a gas-tight exhaust line at the

pump outlet if necessary. Always vent ex-

haust gases appropriately (e.g., into a fume

hood).

+

Never block the gas outlet. The exhaust

line must always be free of obstructions (no

back pressure) to ensure an unimpeded

discharge of gas. The cross-section of the

outlet tubing must be at least the size of the

pump’s exhaust connection.

• Reduce the transmission of vibration. Prevent mechan-

ical load due to rigid pipelines. Insert elastic hoses or

flexible elements as couplings between the pump and

rigid pipes.

Secure hose connections at the pump ap-

propriately, e.g., with hose clamps, to protect

against accidental detachment.

To reduce pump noise emanating from the

pump exhaust port, connect an exhaust hose

or use a silencer (see “Accessories / Spare

parts”, pg. 43).

NOTICE

Advertising
This manual is related to the following products: