In case of overload, the motor is shut down by

a

self-hold thermal cutout in the winding.

Note: Only manual reset is possible. Switch

off the pump and disconnect from the power

source. Identify and eliminate the cause of fail-

ure. Allow the pump to cool down sufficiently

before restart.

°C / °F

I ➨ O

!

3.

2.

1.

IN

Avoid overheating (e.g., due to hot process

gases). Maximum permissible temperature

range: see ”Gas inlet temperatures”, pg. 26.

Make sure the fan’s air supply is adequate. Do

not place the pump on soft surfaces (e.g., rub-

ber foam) during operation. This may reduce

the air circulation in the pump housing and af-

fect the fan’s cooling effect. Check fan regu-

larly for dust/dirt. Clean fan guard grill if neces-

sary to avoid a reduction of ventilation.

air

NOTICE

NOTICE

NOTICE

A warm up period (approximately 15 min.) is

required to ensure that the rated ultimate vac-

uum and pumping speed are attained.

O

➨ I

m

3

/h

Torr/mbar

~15min

Important notes regarding the use of gas ballast

Gas ballast is a continuous purge to keep the pump’s in-

terior as clean as possible and to reduce the possibility of

condensation inside the pump.