Notice – VACUUBRAND MV 10C NT + EK User Manual

Page 32

Advertising
background image

page 32 of 92

• Reduce the transmission of vibration. Pre-

vent mechanical load due to rigid pipelines.

Insert elastic hoses or flexible elements

as couplings between the pump and rigid

pipes.

Note: Flexible elements will compress or

flatten when evacuated if not designed for

use under vacuum.

• Hose connections at the pump inlet must al-

ways be gas tight.

Use a suitable adapter to connect a vacuum

hose (caoutchouc) 15 mm ID (see “Accesso-

ries / Spare parts”, pg. 43), if necessary.

inlet

Use connecting hoses with large diameter and

keep them as short as possible to avoid flow

losses. Locate the pump as closely as possi-

ble to the application.

Secure hose connections at the pump ap-

propriately, e.g., with hose clamps, to protect

against accidental detachment.

Use a suitable valve (see “Accessories / Spare

parts”, pg. 43) to isolate the pump from the

vacuum application. This is to allow the pump

to warm up before pumping condensable va-

pors and to clean the pump after use before it

is switched off.

NOTICE

NOTICE

NOTICE

Advertising
This manual is related to the following products: