6. Advanced Functions

4.

How to use the L358 as an incident illuminance (LUX or FC) meter

1.

Turn the Lumisphere Retracting ring q to lower it to
the

mark position.

2.

Make sure that any compensation (see page 28) is
canceled (

).

3.

Set the meter to EV mode (DIP switch 1) and ISO 100.

4.

Place meter parallel to the light source and take a
measurement.

5.

Convert the measured EV with a conversion table to
find the brightness level.

EV value Lux conversion table

-2
-1

0
1
2
3
4
5
6
7
8

0.63

1.3
2.5
5.0

10
20
40
80

160
320
640

0.88

1.8
3.5
7.1

14
28
57

110
230
450
910

9

10
11
12
13
14
15
16
17
18
19

1300
2600
5100

10000
20000
41000
82000

160000
330000
660000

1300000

1800
3600
7200

14000
29000
58000

120000
230000
460000
930000

1900000

EV

EV

Decimal places

0

0.5

0

0.5

Decimal places

EV value Foot candle (FC) conversion table

-2
-1

0
1
2
3
4
5
6
7
8

0.06
0.12
0.23
0.46
0.93

1.9
3.7
7.4

15
30
59

0.08
0.16
0.33
0.66

1.3
2.6
5.3

11
21
42
84

9

10
11
12
13
14
15
16
17
18
19

120
240
480
950

1900
3800
7600

15000
30000
61000

120000

170
340
670

1300
2700
5400

11000
22000
43000
86000

170000

EV

0

0.5

0

0.5

Decimal places

EV

Decimal places