Appendix a: interference table – ENMET EN Series User Manual

Page 10

Advertising
background image

Sensor/Transmitter, Toxic Gas & Oxygen

ENMET Corporation

8


Appendix A: Interference Table

For Selected Sensors

Ammonia (3E-100 sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

ethanol

1000

1

isopropanol

450

1

hydrogen sulfide

14

10

methanol

1200

3

hydrogen

1000

80

carbon monoxide

300

100

sulfur dioxide

25

-21

nitrogen dioxide

50

25

hydrogen cyanide

10

-18

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

chlorine

5

unsaturated hc (ethylene)

1.0 %

saturated hydrocarbons

abundant

Carbon Monoxide (3E sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

hydrogen

1000

450

nitric oxide

100

25

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

ammonia

100

chlorine

5

hydrogen cyanide

10

ethylene

2.0 % *

carbon dioxide

5,000

methane

10,000

sulfur dioxide

10 *

hydrogen sulfide

10 *

nitrogen dioxide

10 *

isopropanol

1,025 *

gasoline vapor

saturated *

* With onboard filter; continuous high level exposure may reduce the filter efficiency