ENMET EN Series User Manual

Page 14

Advertising
background image

Sensor/Transmitter, Toxic Gas & Oxygen

ENMET Corporation

12


Fluorine (3E 10 sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

chlorine

1

2

hydrogen sulfide

10

-1

sulfur dioxide

5

-3.2

nitrogen dioxide

5

.5

bromine

1

3

ammonia

1000

-3.1

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

carbon monoxide

300

carbon dioxide

100000

nitrogen

100%

hydrogen chloride

10

hydrocarbons

% range

hydrogen

1000

hydrogen cyanide

10

ethanol

4%

Hydrogen (2E 2000 sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

carbon monoxide

50

0

ethylene

500

0

sulfur dioxide

2

0

isopropanol

1090

180

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

ammonia

100

carbon dioxide

1000

chlorine

5

hydrogen cyanide

10

hydrogen sulfide

10*

methane

10000

nitrogen dioxide

10

sulfur dioxide

2

*With onboard filter; continuous high level exposure may reduce the filter efficiency