Table 1: electrochemical sensors – ENMET EN Series User Manual

Page 7

Advertising
background image

Sensor/Transmitter, Toxic Gas & Oxygen

ENMET Corporation

5


4.0

Standard Ranges, Calibration Gases, Voltage Readings at Calibration Points and

Approximate Times to Sensor Stabilization

Table 1: Electrochemical Sensors

Gas

Symbol

Range*

(ppm)

Calibration Gas

Correlation

Reading

(mV)

Stabilize

(min.)

ENMET Gas

Part Number

Ammonia

NH

3

0 – 100

25ppm NH

3

200

4

03218-025

Arsine (2)

AsH

3

0 – 0.5

0.5ppm PH

3

0.375ppmAsH

3

400

3

03202-005

Boron Trichloride

BCl

3

0 – 30

10ppm HCl

10ppm BCl

3

233

3

03202-010

Carbon Monoxide

CO

0 – 100

50ppm CO

300

2

03219-050

Chlorine (2)

Cl

2

0 – 10

5ppm Cl

2

300

4

03231-005

Chlorine Dioxide

ClO

2

0 – 1.0

0.3ppm ClO

2

**

220

4

**

Dichlorosilane

SiH

2

Cl

2

0 – 30

10ppm HCl

3ppm SiH

2

Cl

2

140

3

03202-010

Ethylene Oxide

C

2

H

4

O

0 – 30

5ppm C

2

H

4

O

167

3

03232-005

Fluorine (2)

F

2

0 – 10

3ppm Cl

2

7.5ppm F

2

400

3

03231-003

Hydrogen

H

2

0 – 1000

800ppm H

2

420

3

03227-800

Hydrogen Bromide

HBr

0 – 30

5ppm HCl

25ppm HBr

433

3

03202-005

Hydrogen Chloride (2) HCl

0 – 30

10ppm HCl

233

3

03202-010

Hydrogen Cyanide

HCN

0 – 30

10ppm HCN

233

2

03203-010

Hydrogen Fluoride (1)

HF

0 – 10

5ppm Cl

2

3ppm HF

220

4

03202-005

Hydrogen Sulfide

H

2

S

0 – 100

20ppm H

2

S

180

2

03214-020

Nitrogen Dioxide

NO

2

0 – 30

5ppm NO

2

167

2

03233-005

Nitric Oxide

NO

0 – 100

25ppm NO

200

1.5

03234-025

Oxygen

O

2

0 – 30 %

20.9% O

2

379

2

03296-209

Ozone (1)

O

3

0 – 1.0

0.3ppm O

3

**

220

3

**

Phosgene (1)

COCl

2

0 – 1.0

0.66ppm COCl

2

**

364

4

**

Phosphine (2)

PH

3

0 – 1.0

0.5ppm PH

3

300

1.5

03205-005

Silane

SiH

4

0 – 30

5 ppm SiH

4

167

2

03306-005

Silicon Tetrachloride

SiCl

4

0 – 30

5ppm HCl

25ppm SiCl

4

433

3

03202-005

Sulfur Dioxide

SO

2

0 – 30

10ppm SO

2

233

1.5

03215-010

N

OTE

:

All specifications stated in this manual may change without notice.

(1) Require use of Sample Draw System.
(2) Recommended use of Sample Draw System, for enhanced sensitivity and speed of response.

*0 = 4mA = 100mV, upper point = 20mA = 500 mV

**Gas not available in cylinders, a gas generator must be used

N

OTE

: If a non-standard range and/or a non-standard calibration gas are used, utilize the equations in Section 3.0 to determine

voltage readings.