Sensor/Transmitter, Toxic Gas & Oxygen

ENMET Corporation

9


Chlorine (3E sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

hydrogen sulfide

10

-0.3

sulfur dioxide

5

-1.2

nitrogen dioxide

5

0.1

bromine

1

1.0

chlorine dioxide

0.32

0.3

ammonia

1,000

-1.1

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

carbon monoxide

300

carbon dioxide

100,000

nitrogen

100. %

hydrogen chloride

20

hydrocarbons, general

% range

hydrogen

1,000

ethanol

6.6 %

ammonia

65

Hydrogen Sulfide (3E sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

carbon monoxide

100

3

chlorine

20

-1

ethylene

500

2

hydrogen

100

5

hydrogen

20,000

100

hydrogen cyanide

10

1

sulfur dioxide

10

3

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

ammonia

100

carbon dioxide

5,000

methane

10,000

sulfur dioxide

3

nitrogen dioxide

10