ENMET EN Series User Manual

Page 13

Advertising
background image

Sensor/Transmitter, Toxic Gas & Oxygen

ENMET Corporation

11


Hydrogen Chloride(3E sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant
Concentration in ppm

Instrument Reading
in ppm

hydrogen bromide

5

5

hydrogen sulfide

14

30

sulfur dioxide

5

3.5

chlorine

5

1

hydrogen cyanide

14

1

arsine

330 ppb

.4

ethanol

6.6%

6

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

carbon monoxide

300

carbon dioxide

5000

nitrogen

100%

hydrocarbons

% range

hydrogen

1000

phosgene

.5

chlorinated hydrocarbons

% range

ammonia

300

Hydrogen Fluoride (3E 10 sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

chlorine or bromine

1

.51

sulfur dioxide

2

2

hydrogen chloride

5

.75

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

alcohols (i.e. IPA)

1000

ammonia

100

hydrogen

1000

carbon monoxide

50

carbon dioxide

10%

unsaturated HC(ethylene)

1%

hydrogen sulfide

20

Advertising