ENMET EN Series User Manual

Page 16

Advertising
background image

Sensor/Transmitter, Toxic Gas & Oxygen

ENMET Corporation

14


Ozone (3E 1 sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

chlorine

1

0.80

fluorine

100ppb

0.07

germane

1

1

nitrogen dioxide

1

0.09

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

hydrogen sulfide

1

sulfur dioxide

2

phosphine

100 ppb

carbon monoxide

300

carbon dioxide

5000

nitrogen

100%

nitrogen monoxide

10

hydrocarbons

% range

hydrogen

1000

hydrogen cyanide

10

ammonia

10

hydrogen fluoride

3.8

hydrochloric acid

10

Arsine (2E sensor)

The following gases cause a reading:

Interferant
Gas/Vapor

Interferant Concentration
in ppm

Instrument Reading
in ppm

phosphine

100 ppb

110

chlorine

5

-400

hydrogen cyanide

10

100

ammonia

100

10

diborane

100

35

silane

1

100

germane

1.1

100

hydrogen selenide

50 ppb

5

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

hydrogen sulfide

1*

sulfur dioxide

2

hydrochloric acid

5

carbon monoxide

300

carbon dioxide

5000

nitrogen

100 %

hydrocarbons

% range

hydrogen

1000

*With onboard filter; continuous high level exposure may reduce the filter efficiency