ENMET EN Series User Manual
Page 15
Sensor/Transmitter, Toxic Gas & Oxygen
ENMET Corporation
13
Nitric Oxide (3E 100 sensor)
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
hydrogen
1000
carbon monoxide
1000
carbon dioxide
10000
sulfur dioxide
50
chlorine
5
nitrogen dioxide
100
saturated HC, alcohols
abundant
Nitrogen Dioxide (3E sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
chlorine
1
3
ozone
1
0.7
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
hydrogen chloride
5
nitrogen oxide
100
hydrogen
1000
carbon monoxide
1000
carbon dioxide
10000
saturated HC, alcohols
abundant
sulfur dioxide
50
ammonia
30